Operation
▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬
SafeCheck 200 Operator’s Manual
17
During the calibration procedure it also shows you the level of gas that the
sensor is detecting. At anytime during the calibration procedure, calibra-
tion can be terminated by depressing any key. At the conclusion of a suc-
cessful calibration for a gas, depress the I/O key or the ▲ arrow key to
return to the sensor list screen.
Repeat the procedure, as necessary, for any other gases being measured.
To return to the main CAL. menu, press the ▲ arrow key.
Operation
▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬
SafeCheck 200 Operator’s Manual
18
• All Cal:
This operates the same as INDIVIDUAL CAL except that it steps auto-
matically through all of the installed sensors to
calibrate all of them. The unit first waits to do a
CALIBRATION ZERO. This is automatically
done before any calibration to adjust for drift in each sensors' sensitivity
which occurs over time and after use.
NOTE: Fresh uncontaminated air is necessary for the proper outcome of
this procedure. (Zero the unit away from heavy traffic, excessive cigarette
smoke, and industrial fumes.) If such an atmosphere is difficult to reach,
zero air cylinders are available.
Press the I/O key when you are sure that you are in a clean air environ-
ment. The unit goes through and automatically zeros all the installed sen-
sors.
It now instructs you to apply the gas to the unit.
Apply the calibration gas for sensor #1 in the
unit and press the I/O key. It is very important at
this point if an O
2
sensor is installed and the
calibration level used is some thing other than 20.9% that you apply the
calibration gas before pressing the I/O key. Not adhering to this procedure
in this case will result in an incorrect calibration for O
2
.
At this point the unit takes over and runs the calibration. It asks that you
apply gas for the first sensor it finds. On the display, the unit asks you to
apply the concentration of calibration gas that was specified in SPAN
ADJUST, showing the expected value and displaying the message
"Waiting". From here on, it operates just like INDIVIDUAL CAL and
steps through each gas in turn asking you to apply each gases' cal. level
until it is done or you abort the process.
Calibration Zero
Ready? ■
Apply Gas …
Enter to Cal.